The turbidity meters Turb® 430 IR acc. to DIN ISO 27027 and Turb® 430 T acc. to US EPA 180.1 meet the requirements of the nephelometric measurement at 90° stray light for drinking water with highest precision, especially in the range < 1 NTU. The measuring range comprises 0-1100 NTU/FNU with automatic measuring range switching and is therefore suitable for water analysis as well as quality control, such as gasoline, cell cultures, as well as in the food and beverage industries.
Features:
- Lab precision, suitable for drinking water
- Adjustable calibration intervals and GLP-compliant documentation
- Stray light behaviour according to pharmacopoeia 5.0
- The measurement and automatic 3-point calibration runs self-explanatory and menu-controlled
- The calibration takes place via an AMCO Clear®standard set (0.02-10-1000 NTU). Up to 1000 data sets including sample ID numbers can be stored and later put out GLP-compliant via the PC software LSdata
- The quality assurance of the measured results is supported by calibration interval setting with documentation.
- This makes Turb®430 not only a portable meter – also available in a practical field set – but also a precise and reliable benchtop meter for applications of up to 1100 NTU/FNU with optimized data administration