Noxious Gas Test
To prove the resistance of the technical products to corrosive gases, Weiss developed the corrosion test chamber cabinets AirVent. The AirVent enables an exact dosage of the corrosive gases with a climate conditioned air volume. The main constituents of the corrosive, atmospheric trace elements are: sulphur dioxide (SO2), nitrogen oxide (NOx), hydrogen sulphide (H2S) and chlorine gas (CL2).
Features
Features Corrosive Gas Test System
- Gas dosing with thermal mass flow controller, concentration ranges from ppb to ppm
- Inner test cabinet with connection for gas sampling and entry port NW 50mm
- Air exchange rate in inner test cabinet can be set between one and four per hour
- Door with window included in basic equipment
- Specimen can be observed (inner test cabinet is made of acrylic glass (PMMA))
- Up to four gasses can be dosed simultaneously
- Inner test cabinet, shelf and door are made of acrylic glass
- Pressure monitoring between inner test cabinet and outside as a safety feature
- Compliance with test standards such as IEC 60068-2-60, IEC 60068-2-42/43, ISO 21207 and others
Test gases
- H2S
- SO2
- Cl2 with carrier gas N2 (nitrogen)
- NO2, with carrier gas synthetic air
- other test gases on request